Tube No.
0
30
40
50
20
10
Ethanol〔ppm〕?
FIG.1 Influence of Ethanol
Ethyl cellosolve indicating value 〔ppm〕
100 200 300 400 500
(300mR)? (100mR)?
0
300
400
500
200
100
1,1,1-Trichloroethane〔ppm〕?
FIG.2 Influence of 1,1,1-Trichloroethane
Ethyl cellosolve indicating value 〔ppm〕
100 200 300 400 500 600
(300mR)?
1. PERFORMANCE
1) Measuring range :5-500 ppm
Number of pump strokes 3(300mr)
2) Sampling time :4.5 minutes/3 pump strokes
3) Detectable limit :2 ppm
4) Shelf life :2 years
5) Operating temperature :10~35℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 3 pump strokes
8) Colour change :Yellow→Pale blue
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
Chromium oxide is reduced.
C2H5OCH2CH2OH+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
500
400
500
200
150
100
50
20
5
300ml
ppm
(White)
DEHUMIDIFIER
??
????
190U ETHYL CELLOSOLVE
Substance Interference Coexistence
Alcohols FIG.1 Similar stain is produced. Higher readings are given.
Ethers 〃〃
(Pmaroarfefi nth haynd Cro3c)arbons Whole reagent is changed to Brown. 〃
Aromatic hydrocarbons 〃〃
Esters 〃〃
Ketones 〃〃
Halogenated hydrocarbons FIG.2 〃〃
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 10℃ 15℃ 20℃ 25℃ 30℃ 35℃
(ppm) (50°F) (59°F) (68°F) (77°F) (80°F) (95°F)
500 800 620 500 410 335 265
400 620 490 400 325 260 200
300 450 365 300 245 195 145
200 290 245 200 160 125 95
150 215 185 150 120 90 70
100 145 125 100 80 60 45
50 75 65 50 40 30 25
20 30 25 20 15 12 10
5 10 7 5 4 3 2